This manual describes the update process of the software running on the sensor, also called DFU (device firmware update). The manual applies to c-med° alpha as well as cosinuss° Two sensors.
Please note: To switch to a different Bluetooth Mode, please see the instructions here.
Please note: Additional information about updating the cosinuss° Two with the Customer Flashing Tool is available here.
Open the cosinuss° LabApp. You will be asked to enter the sensor's serial number which can be found on the sensor itself marked with SN. When you start the app the first time, it will ask for permissions to access Bluetooth and location (Android). Both are necessary for connecting the sensor.
Next, the app will connect to the sensor and navigate to the vital signs screen. If the app says “unable to verify” your phone probably has no internet connection.
By default your sensor's Bluetooth Mode is set to Broadcast Vitals, please change to Basic Vitals before performing the actual DFU:
To start the DFU:
Tap on the notification “Pair & connect” and enter the code 111111 (6 times the 1), confirm with “OK” and then press “UPDATE” to start the update process. “Updating… 2%” informs you about the DFU's progress, this will take a few minutes. The DFU continues even with phone's screen turned off. During the DFU, the sensor turns it's LEDs off.
The speed of the update will vary depending on the strength of the connection and the smartphone's Bluetooth capabilities. If everything completes successfully, the device will reset itself with the new firmware.
By default your sensor's Bluetooth Mode should be set to Broadcast Vitals, so please change back with:
Finally do not forget to disconnect your sensor in the app, otherwise your sensor is not available for other apps or applications.
Possible reasons that the pairing notification is not showing up: