Remote vital signs monitoring using the c-med° alpha and the Gateway.
Model numbers: c-med° alpha: MS01 Gateway: LG01
This product is manufactured by Cosinuss GmbH, Kistlerhofstraße 60, D-81379 Munich https://www.cosinuss.com
For health questions and medical advice, please always consult your doctor. Cosinuss GmbH does not issue any medical recommendations. Cosinuss GmbH is not liable for any abusive or unauthorized use of the product. Cosinuss GmbH is not liable for the loss or misuse of personal data or the product. There are no replaceable parts in the device. Supplied components must not be exchanged. The components must never be exchanged, or the device manipulated without the consent of Cosinuss GmbH.
View the list of signs and symbols used within the labeling and packaging context of the cosinuss° Health Platform and its in-ear sensors:
Please note, the first prototype version of the c-med used a different label, more information about this prior label version can be found here.
Part number | Identifier | Length | Description |
---|---|---|---|
GTIN (country prefix, manufacturer number, article number, check digit) | (1) | 13+1 | Contains information regarding the product and manufacture (+country) to identify the product worldwide |
Manufacturing Date | (11) | 6 | Format: yymmdd |
Type | (10) | 4 | Model of sensor device (e.g.: MS01) |
LOT | (10) | 4 | Identification to track production batch |
Serial number | (21) | 6 | Unique serial number |
Number | Definition |
---|---|
1 | Adjustable sensor neck |
2 | Red/Infrared LED |
3 | Contact thermometer |
4 | Infrared thermometer |
5 | Photodiode |
6 | Antitragus curve |
7 | Pickaxe |
8 | Status LED |
9 | Circuit board and battery |
10 | Charging LED |
11 | Charging contacts |
1 x USB charging cable
Number | Definition |
---|---|
1 | Micro SD-Card Slot |
2 | Power LED |
3 | Power supply socket |
4 | Status LED |
5 | HDMI Socket |
6 | AUX Socket |
7 | Ethernet Socket |
8 | USB Sockets |
(9) bottom | Labeling |
1 x Power supply, 1 x Modem Stick
Number | Definition |
---|---|
1 | USB Connector |
2 | Status LED |
This diagram shows how the Internet stick is connected to the Gateway:
Do not try to remove the SIM stick, as it is fixed on the Gateway device.
The SIM cards are activated automatically as soon as you connect the Gateway with a power source. The SIM stick will then blink for about 1–2 minutes until it is activated and ready for use.
Now, the Gateway will use the LAN connection to connect to the internet.
If a connection is not possible, make sure that your network permits the connection with external devices. If you need to register the MAC address of the Gateway, you find the MAC address in your cosinuss° Health account by going to the tab 'RECEIVERS' and clicking on the serial of the Gateway that you use. In the list of attributes that appears, you will find the value for the MAC address. In a clinical setting, it can occur that a connection with a LAN cable is not possible due to security settings and thus no Internet connection can be established for the Gateway. In these cases, use an Internet connection via an Internet stick or WLAN.
How to connect the gateway to Wi-Fi is described in more detail here.
Our in-ear sensors are already partially charged when delivered and should be equipped with enough battery power for a quick product test. In some cases, the battery of the sensor may be no longer charged and needs to be recharged before use. For prolonged use of the device and to ensure maximum battery performance, the device should be fully charged.
Charging the c-med° alpha:
The charging LED light of the c-med° alpha lights up red when the sensor has been inserted into the charging case. Charging time for full battery capacity is approx. 1 hour. The red light will turn off when the battery is fully charged. Afterward, the sensor can be unplugged.
When you place the sensor back into the charger box, first push the bottom of the sensor under the small overhead part of the charger box and only then push the entire device into the charger box. This notice is additionally attached to the charging box of the c-med° alpha. You can find this instructional label displayed here below.
To conserve power, advertising will occur periodically every 60 seconds in short intervals of 10 seconds. When the blue LED is blinking rapidly, the c-med° alpha is advertising. If no LEDs are blinking, the sensor is off and may need to be recharged.
It is crucial to ensure a firm and comfortable fit of the sensor in the ear so that the most accurate values can be measured. The sensor needs to be inside the ear to start a measurement!
Step-by-step application of the cosinuss c-med° alpha:
During an active recording both, the c-med° alpha and the gateway status LEDs, are blinking green.
As soon as you take the c-med° alpha from the charging box, the device will boot up and begin advertising. How the sensor can be paired with other devices using Bluetooth is predetermined by the selected mode.
The c-med° alpha can be used in two different settings:
In order to record data, you need to add at least one receiving device (the Gateway) and one sensor device (c-med° alpha). In addition, a user account needs to be assigned to a cosinuss° sensor and the Gateway. If you want to adjust this assignment, please refer to the following bullet points for more information. In some cases, we have already registered all your devices and created the corresponding user accounts. In all cases, however, make sure that users have already been created and that the corresponding devices have also been assigned to them.
A subject needs to be assigned to a Gateway and a c-med° alpha sensor. Moreover, a project needs to be connected with a subject. Therefore, please follow the instructions below:
In some cases, it might be necessary to reallocate sensors and Gateway or change subject information retrospectively. For example, if a different sensor size is needed or a subject is no longer part of a study. For this, select PEOPLE in the menu bar and click on the pseudonym, which is displayed in blue font and also known as the hash ID of the person. Then you enter the person detailed view. Here, more details about one person are listed, and subsequent adjustments can be made. This is the same dialog box that opens automatically after a new person is created, so all the settings described above can also be edited retrospectively. Additional information about this can be found here.
The available and displayed setting options depend on what user rights a user possesses. You can find an overview of the different user rights here and how to set up a project.
After selecting PEOPLE in the menu bar, the following options appear when you hover over the row of one particular user.
Further information about these edit settings is provided here.
For more information about the CSV data download, click here.
Registration of the c-med° alpha and the Gateway:
In order to record data, you need to add at least one receiving device (Gateway) and one sensor device (c-med° alpha). Please assign one person to one sensor to one gateway at a time for the most reliable data acquisition and transfer to the server.
In order to see the vital signs data inside cosinuss° Health make sure that:
The vital signs data is available for each pseudonym (so for each registered person). By default, you see the results of your last three minutes measurement. A detailed explanation of display options and further settings of the data view is available here.
Numeric legend | Displayed information |
---|---|
1 | Patient pseudonym |
2 | Serial number of sensor |
3 | Battery level |
4 | Time of the last measurement |
5 | Deviation score |
6 | Vital signs parameters |
7 | Quality indicators |
The current standard configuration of the cosinuss° Health Platform is to start measurement of 3 minutes duration every 15 minutes. After the measurement is completed, the respective values shown above are calculated and shown on the Dash. A detailed description of each parameter can be found here and the method behind the Deviation Score calculation here.
! Due to the currently used standard configuration of the cosinuss° Health Platform, it might take a few minutes until values will be visible on the Dash.
For the most accurate data measurement, ensure that there is as little body movement as possible during the measurement. In particular, talking and chewing can significantly distort the measurements. Also, incorrect placement of the sensor inside the ear canal and disturbed data transmission between Gateway and sensor can result in inaccurate measurements. So, please make sure that the sensor is inserted as described and a stable connection between the sensor and the Gateway is provided. In addition, the quality score of the measured data indicates if the acquired data is reliable (qs>=40) or not (qs<40).
Note: Establishing the connection to start the measurement may take a few minutes. If a problem occurs with your c-med° alpha, please follow this troubleshooting information first. If this does not solve the problem, please contact cosinuss° customer service. See chapter: Customer service.
Problem | Indicator | Reason | Solution |
---|---|---|---|
The c-med° alpha does not switch ON. | Status LED does not turn ON. | Battery is empty. | Charge in-ear sensor. |
The c-med° alpha does not turn OFF. | Status LED does not turn OFF. | The sensor is not correctly positioned in the charging box. | Position sensor with good contact of the charging pins. |
The c-med° alpha is not charging. | Charging LED does not turn ON. | The sensor is not correctly positioned in the charging box. | Position sensor with good contact of the charging pins. |
Problem | Indicator | Reason | Solution |
---|---|---|---|
The c-med° alpha does not connect to the app. | The sensor stays in advertising mode. Status LED blinks blue. | The Bluetooth function of the smartphone is turned OFF. | Turn ON Bluetooth on the smartphone and retry. |
The Bluetooth connection is interrupted repeatedly. | The status bar in the monitoring screen shows varying Bluetooth states. | See chapter: Warnings and precautions >Bluetooth | See chapter:Warnings and precautions >Bluetooth |
The serial number is wrong. | Error message. | Typing error. | Correct typing error and try again. |
The serial number is not recognized. | Error message. | The smartphone has no network connection. | Establish network connection and retry. |
The serial number is not valid. | Error message. | No match on server. | Contact customer support. See chapter: Customer support |
Problem | Indicator | Reason | Solution |
---|---|---|---|
No vital signs are displayed. | The value shows no value (–.-) | No in-ear sensor connected. | Connect your sensor. |
No in-ear sensor in Bluetooth range. | See chapter:Warnings and precautions >Bluetooth | ||
No new values incoming since 30 seconds. | Check Bluetooth connection. |
||
Measured values are out of range. | See chapter: Product specifications |
||
Vital signs are marked invalid. | Question mark in vital signs card. | Quality index and/or Perfusion index are low. | Minimize influencing factors. See chapter: Warnings and precautions |
Problem | Indicator | Reason | Solution |
---|---|---|---|
The in-ear sensor hurts. | Individual perception. | Wrong fitting. | Check the correct fitting of the in-ear sensor. See chapter: Apply the in-ear sensor to the ear |
Wrong in-ear sensor size. | Check the long term behavior of the quality indices. |
||
Exceeded wearing time. More than 12 hours without interruption. | Cancel the measurement. See chapter: Warnings and precautions |
||
Sensor mechanically broken. | Cancel the measurement. See chapter: Warnings and precautions |
||
The monitored person’s ear is injured. | Cancel the measurement. See chapter: Warnings and precautions |
Status LED | Meaning |
---|---|
Green, solid | Everything’s fine. Measurement paused. Gateway ready. No measurement, no upload, no update is taking place. Wait for the next measuring interval. |
Green, blinking | Everything’s fine. Active recording/measurement in extended data mode or ongoing broadcast vitals transmission. Do not unplug the Gateway! |
Red, solid | Error: server sync failed The Gateway could not sync with the server. (e.g. problems with the Internet connection). Check internet connection. |
Blue, blinking | Everything's fine. Gateway update in progress. Do not unplug the Gateway! |
Blue, pulsating | Everything's fine. Ongoing WiFi configuration process with Cosinuss Connect App. Do not unplug the Gateway! |
Blue, solid | Not synced yet. No successful server sync has occurred yet. |
Blue-green, alternating | Everything's fine. Files are being stored and updated to the server. Do not unplug the Gateway! |
Applies to Gateway versions 2.7.x and higher.
If no troubleshooting is successful, the system requires a RESTART. Do this by plugging the Gateway OFF and back ON to the socket. Try to avoid restarts, if the status LED is blinking or pulsating.
Color | Behavior | Definition |
---|---|---|
GREEN | blinks 2x every 2 sec. | The USB Modem stick is switched on. |
GREEN | blinks 1x every 2 sec. | The USB Modem stick is registered in a 2G network. |
BLUE | blinks 1x every 2 sec. | The USB Modem stick is registered in a 3G/3G+ network. |
GREEN | lights up permanently | A data connection to a 2G network is active. |
BLUE | lights up permanently | A data connection with a 3G network is active. |
CYAN | lights up permanently | A data connection with a 3G+ network is active. |
- | LED turned OFF | The USB modem stick is disconnected or is not supplied with power. |
Color | Behavior | Definition |
---|---|---|
RED | solid | The USB Modem stick is switched on. |
GREEN | blinking | Register and connect to 2G/3G |
GREEN | solid | A data connection with a 2G/3G network is active. |
BLUE | blinking | Register and connect to 4G |
BLUE | solid | A data connection with a 4G network is active. |
- | LED turned OFF | The USB modem stick is disconnected or is not supplied with power. |
LED behavior | Description |
---|---|
Blinking blue (a short blink every 2.5 seconds) | Advertising. The in-ear sensor is broadcasting data to available receiving devices and is ready to establish a connection. |
Blinking green (a short blink every 2.5 seconds) | Connected. The in-ear sensor is working properly. |
Blinking red (a short blink every 2.5 seconds) | Connected. The in-ear sensor is working properly, but the battery is low. |
Solid red (only applicable when the sensor is placed in the ear) | Connected. But the in-ear sensor is probably damaged, due to improperly working sensors and circuits. |
LED is OFF | In-ear sensor is turned OFF. In-ear sensor is positioned correctly in the charging box. |
LED behavior | Description |
---|---|
Solid red | In-ear sensor is charging. |
LED is OFF | In-ear sensor is fully charged. |
Take good care of your c-med° alpha to ensure long-lasting performance. The sensor head and the sensor lens are the most sensitive parts of the c-med° alpha. They should always be clean and faultless to ensure accurate monitoring. Store and charge your in-ear sensor only within the c-med° alpha charging box. If the sensor head or sensor lens is damaged, please contact customer service. See chapter: Customer service.
Before and after each use clean the in-ear sensor as shown in the following instructions. For disinfection and cleaning Cosinuss GmbH recommends the use of ‘Schülke mikrozid® AF liquid’ as an alcohol-based cleaning agent. Alternatively, any other disinfectant with the following composition can be used: 35% (w/w) 1-propanol, 25% (w/w) ethanol, no fragrances, undiluted application.